适用环境:Python 3.10+ and uv; pinned workflows use NeuroKit2 0.2.13. Core processing needs NumPy, SciPy, pandas<3, scikit-learn, matplotlib, PyWavelets, requests, and setuptools; selected EEG, cvxEDA, plotting, file-format, and RQA features need separately locked optional packages.

输入资料
获准使用的生理信号与采样信息 研究目标和 NeuroKit2 环境 阅读本节完整说明
交付内容
信号处理与特征资料 研究分析和质量检查记录 阅读本节完整说明
环境与依赖
For optional features, create a uv project, add only the packages actually required at reviewed exact versions, and commit/review the resulting uv.lock before uv sync --locked. NeuroKit2 exposes an upstream full extra, b… 阅读本节完整说明
  1. 获取完整包,保留配套文件的目录结构。
  2. 按照原文准备所需工具、依赖与输入资料。
  3. 阅读对应使用步骤与输入输出说明,在自己的环境中验证结果。

使用提示摘自原文,配套文件保留来源版本;实际运行条件以各包说明为准。

正文

获取原文件

references/ppg.md

Checked 2026-10-01 against NeuroKit2 0.2.13 stable runtime/source, the official PPG API, 0.2.13 release notes, and measurement guidance.

Acquisition contract

Record sensor mode (reflectance/transmission), wavelength(s), anatomical site, attachment/contact pressure, device/firmware, raw unit/range, ambient-light handling, sampling rate, clock, temperature/perfusion, activity/posture, and motion/accelerometer channels. Validate across representative skin pigmentation, anatomy, age, vascular state, motion, and intended population.

PPG is optical blood-volume-pulse measurement, not cardiac electrical activity. Pulse timing and morphology depend on site and vascular/transit dynamics.

Stable high-level pipeline

signals, info = nk.ppg_process(
    ppg,
    sampling_rate=100,
    method="elgendi",
    method_quality="templatematch",
)

Pinned 0.2.13 default columns:

PPG_Raw, PPG_Clean, PPG_Rate, PPG_Quality, PPG_Peaks

info contained PPG_Peaks, sampling_rate, and peak/correction method metadata. The live doc's older codebook can omit PPG_Quality; stable runtime/source includes it.

Cleaning and peak methods

For explicit selection:

clean = nk.ppg_clean(
    ppg,
    sampling_rate=100,
    method="elgendi",
)
markers, peak_info = nk.ppg_peaks(
    clean,
    sampling_rate=100,
    method="elgendi",
    correct_artifacts=False,
)

Stable cleaning methods include:

  • elgendi;
  • nabian2018 (can use expected heart rate);
  • langevin2021;
  • goda2024; and
  • none.

Stable peak methods include:

  • elgendi;
  • bishop (peaks plus pulse onsets);
  • charlton (MSPTDfast v2; peaks plus onsets); and
  • charlton2024 (superseded v1).

Method-dependent extra outputs are a primary reason not to hard-code one schema. Validate peak/onset performance against labeled data at the actual site, rate, perfusion, motion, and population.

correct_artifacts=True uses the cardiac peak-correction path. Retain raw/corrected peaks and correction categories; correction cannot repair a low-quality optical waveform.

Quality outputs added/expanded in 0.2.13

quality = nk.ppg_quality(
    clean,
    peaks=peak_info["PPG_Peaks"],
    sampling_rate=100,
    method="templatematch",
)

Stable quality methods and scales differ:

  • templatematch: continuous similarity, typically 0–1;
  • dissimilarity: unbounded, where zero is highest similarity;
  • ho2025/interval-consistency path: binary interval quality;
  • skewness, kurtosis, entropy: unbounded windowed metrics;
  • perfusion: AC peak-to-peak / absolute raw DC mean × 100; requires raw PPG and is not bounded at 100 in the implementation; and
  • relative_power: 0–1, requires raw PPG, and defaults to 60 s windows.

No threshold is universal across these outputs. Name the method and direction/scale. For perfusion, a near-zero DC denominator can explode the ratio; exactly zero returns zero in the implementation. Neither outcome establishes poor/good sensor quality. The docstring's advertised 0–100 range is not enforced by stable source/runtime. Short signals can be invalid for a method's default window. ppg_process() passes peak indices—not the whole info dict—to quality estimation.

Quality relative to an average pulse does not prove physiological accuracy: repeated motion-corrupted beats can be morphologically consistent. Combine morphology, motion, contact/perfusion, missingness, detector agreement, and endpoint-specific validation.

Sampling and preprocessing

Sampling requirements depend on endpoint:

  • pulse rate needs less bandwidth than morphology, onset timing, or derivatives;
  • wrist wearables commonly use lower rates than laboratory finger systems;
  • resampling cannot recover missing onset precision or a dicrotic feature; and
  • a nominal rate does not establish clock accuracy or anti-alias filtering.

Do not prescribe one universal minimum. Validate rate and filters for the exact detector/feature and report native/processed rates. Process at native rate before multimodal alignment when possible.

Motion, contact pressure, ambient light, vasoconstriction, temperature, pigmentation, site, and clipping can all change amplitude/morphology. Preserve a quality/artifact mask and accelerometry where available. Avoid interpolating through corrupted pulses.

PPG-derived variability is PRV

prv = nk.hrv_time(peak_info, sampling_rate=100)

NeuroKit2 accepts PPG peaks in HRV functions, but interpretation remains pulse-rate variability. PRV contains pre-ejection and pulse-transit variability and can differ from ECG HRV by site, posture, respiration, activity, temperature, and vascular state.

For an HRV-equivalence claim:

  1. acquire synchronized ECG and PPG;
  2. validate pulse/beat matching and lag/drift;
  3. prespecify agreement metrics and acceptable error for each HRV endpoint;
  4. test rest, task/activity, motion, and relevant populations/sites; and
  5. report PRV terminology when equivalence is not established.

Do not infer ECG morphology, rhythm diagnosis, oxygen saturation, blood pressure, or arterial stiffness from this basic PPG pipeline.

Event and interval analysis

epochs = nk.epochs_create(
    signals,
    events,
    sampling_rate=100,
    epochs_start=-1,
    epochs_end=10,
    baseline_correction=False,
)
event_features = nk.ppg_eventrelated(epochs)
interval_features = nk.ppg_intervalrelated(signals, sampling_rate=100)

Documented event fields include baseline/min/max/mean/SD rate, times, and polynomial trend coefficients. Interval output includes mean rate and HRV-family columns. Availability depends on input columns, duration, and release; inspect runtime output.

Pulse morphology

ppg_segment() returns a dict of pulse epochs. Morphology comparisons require:

  • consistent site, attachment, pressure, wavelength, and polarity;
  • validated onsets/peaks and quality masks;
  • appropriate baseline/amplitude normalization;
  • sufficient sampling/bandwidth;
  • control of heart rate and vascular state; and
  • endpoint-specific evidence.

A dicrotic feature in a processed waveform does not by itself validate aortic valve timing or arterial stiffness.

Interpretation boundary

Use these tools for research and education. They are not validated here for arrhythmia, oxygen saturation, blood pressure, disease detection, remote patient monitoring, alarms, or wearable medical-device validation.